PFT-500
Carrier-Tape Peel Force Tester
Peel Force Tester JE PFT-500
Feature/Function
1. Our JE PFT-500 can be easily linked to any types of computers.
1
2. Our JE PFT-500 is compatible with WINDOWS 10, XP software and easily
installed.
3. Our JE PFT-500 has following functions:
- Key-in operations: Company name, operator name, file name2, up/down limits of
the peel force, the setting for delay length of data reading
3 ,length of carrier for
testing
4, peel speed
5, the setting for protection limit
6, manual/auto operation mode
7.
- Data Input: cover tape specification, carrier tape specification, sealing conditions,
type of Taping machine, testing environment, etc. and all the parameters input.
- Data Process: maximum Peel Force, minimum Peel Force, range of Peel Force,
average of Peel Force, deviation of testing data.
4.The force gage installed in our machine can be adjusted transversely or angularly, and
the testing sample can be aligned easily.
5. The height adjustment of the testing sample can be suitable for any Ko value of the
carrier Tape.
6. As soon as the software begins to run, the force gage will zero automatically.
7. The testing curve scale can expand tagging along the data of the peel force and the
extension of the measure distance.
9. Our JE PFT-500 supports color printers to produce Chinese/English testing reports and
QC checklists
8.
Remark
Note 1. Equipped with microcomputer control, our machine provides two-way
communication between computers.
Note 2. Data files can be named freely by users, or numbered and dated automatically by
our system.
Note 3. The length for the delay of data reading can be preset to cut the invalid segments
of the curve caused by not strengthening the testing sample initially.
Note 4. The testing length indicates the testing stroke within the distance of 50 to 220
mm, which can be set freely. In the process of testing, the LCD of the machine
can indicate the digital value of the displacement.
Note 5. The peel speed can be set freely within the range of 100 -300 mm per minute,
showing on the LCD of the machine before the testing process.
Note 6.Our system will stop the machine and display warning messages on the monitor,
once the testing data is over the safety value which can be set before the process.
Note 7. Auto mode is suitable for a lot of testing samples; you need only one press to
finish the procedure -- to zero the force gage, to catch and save data, to print
testing reports, to reset the machine, and open a new file in serial number for
next test.
Note 8. Our testing software provides a QC summery report with significant results from
at most 20 different testing samples.